Brookfield Engineering Labs., Inc. Page 4 Manual No. M/03- 65-C0508 C0808 I. INTRODUCTION The Brookfield DV-II+Pro Viscometer measures fluid viscosity at given shear rates.Viscosity is a measure of a fluid’s resistance to flow. A file of Test Results that is saved onto a USB Flash Drive can be viewed on any DV2T Viscometer. Within the View Results function, the user can access the internal memory of the DV2T. Range Figure II-6 TIP: The Range value is the same as the AutoRange available. ©2019 AMETEK.Inc. All rights reserved. By continuing to use this site, you agree to our Policy and Cookie policy. Brookfield Engineering Labs., Inc.Page 3 Manual No. M/98-350-D0902 I. INTRODUCTION The Brookfield DV-E Viscometer measures fluid viscosity at given shear rates. Viscosity is a measure of a fluid’s resistance to flow. You will find a detailed description of the mathematics of. View and Download Brookfield DV-II+ operating instructions manual online. DIGITAL VISCOMETER. Measuring Instruments Brookfield DV2T Operating Instructions Manual. Touch screen viscometer (91 pages). The Brookfield Viscometer/Rheometer will produce an accurate and repeatable torque reading under any measurement circumstance.
Brookfield Engineering Labs., Inc.
Page 76
Manual No. M13-167-B0614
When a UL or DIN UL Adapter is used, the water bath is stabilized at the proper temperature:
1) Put the proper amount of viscosity standard fluid into the UL Tube (refer to the
UL Adapter instruction manual).
2) Attach the spindle (with extension link and coupling nut) onto the DV2T.
3) Attach the tube to the mounting channel.
4) Lower the tube into the water bath reservoir, or if using the ULA-40Y water jacket,
connect the inlet/outlets to the bath external circulating pump.
5) Allow 30 minutes for the viscosity standard, sample chamber and spindle to
reach test temperature.
6) Measure the viscosity and record the Viscometer reading.
Note: The spindle must
rotate at least five (5) times before readings are taken.
Calibration Procedure using a Helipath Stand and T-Bar Spindles
When a Helipath Stand and T-Bar spindles are used:
1) Remove the T-bar spindle and select a standard LV (#1-3) or RV, HA, HB (#1-6)
spindle. Follow the procedures for LV (#1-3) and RV, HA, HB (#1-6) Brookfield
spindles outlined above.
2) T-Bar spindles should not be used for verifying calibration of the DV2T
Viscometer.
Calibration Procedure for Spiral Adapter
Brookfield recommends a two step check. First, verify the calibration of the viscometer
Free voice command download. Voice command free download - Vocal Warm Up by Musicopoulos, FireFly Voice Command Browser, Voice Command Smart Keyboard, and many more programs. Voice command free download - FireFly Voice Command Browser, Voice Command Smart Keyboard, e-Speaking Voice and Speech Recognition, and many more programs. To the maximum extent permitted by applicable law, the aggregate liability of Mega Voice Command and its affiliates, officers, employees, agents, suppliers and licensors, relating to the services will be limited to an amount greater of one dollar or any amounts actually paid in cash by you to Mega Voice Command for the prior one month period.
using the standard viscometer spindles (LV #1-3, RV #2-6, HA #2-6 and HB #2-6 or
cone/plate spindles) as detailed in this appendix. Second, verify the calibration of the
viscometer using the Spiral Adapter. The use of an accessory device may increase the
accuracy of measurement associated with the DV2T.
1) Place the viscosity standard fluid (in the proper container) into the water bath
(refer to the Spiral Adapter instruction manual).
64 bit operating system download. 2) Attach the spindle to the viscometer. Attach chamber (SA-1Y) and clamp to the
viscometer.
3) Lower the DV2T into measurement position. Operate the viscometer at 50 or
60 RPM until the chamber is fully flooded.
4) The viscosity standard fluid, together with the spindle, should be immersed in
the bath for a minimum of 1 hour, stirring the fluid periodically (operate at 50 or
60 RPM periodically), prior to taking measurements.
5) After 1 hour, check the temperature of the viscosity standard fluid with an accurate
thermometer.
6) If the fluid is at test temperature (
Brookfield Dv2t Viscometer Manual Free
+
/- 0.1
°
C of the specified temperature, normally
25
°
Brookfield Dv2t Viscometer Manual 2017
C), measure the viscosity and record the viscometer reading.
Note: The spindle
must rotate at least five (5) times for one minute, whichever is greater before readings are
taken.
7) The viscosity reading should equal the cP value on the viscosity fluid standard
to within the combined accuracies of the viscometer and the standard (as
discussed in the section entitled, Interpretation of Calibration Test Results).

Brookfield Engineering Labs., Inc.
Page 5
Manual No. M13-167-B0614
I. INTRODUCTION
The Brookfield DV-II Viscometer series has been the leading industrial viscometer since it
was first introduced in 1985. Brookfield has continued to develop and improve the DV-II to
provide the best value in the market for both quality control and research customers. The
Brookfield DV2T Viscometer continues in this tradition of innovation, quality and value. The
incorporation of a full color graphical touch screen display has allowed for a new and improved
user interface which preserves the single speed data collection methods while making all of
the advanced features of the DV2T readily available.
The Brookfield DV2T Viscometer measures fluid viscosity at given shear rates. Viscosity is
a measure of a fluid’s resistance to flow. You will find a detailed description of the science of
viscosity in the Brookfield publication “More Solutions to Sticky Problems”, a copy of which
was included with your DV2T.
This manual covers the DV2T Viscometer and the DV2T Extra Viscometer. The DV2T Extra is
a special package that includes several optional features to maximize the utility of the DV2T.
The DV2T Extra includes: Ball Bearing Suspension (not included in LV versions), EZ-Lock
spindle connection, Model Q Lab Stand, and RheocalcT Software. These differences are
detailed in the Components section of the manual. All references to the viscometer in this
manual will be made as DV2T. All of the optional features of the DV2T Extra may be added
individually to the DV2T Viscometer.
• The DV2T offers exceptional versatility in modes of control, allowing choice of traditional
standalone operation and automatic operation through programs downloaded from
the PC or with complete control by PC using Brookfield RheocalcT Software.
• The DV2T can be used as a traditional Brookfield viscometer for collection of single
speed viscosity data through the easy to use touch screen; just select the spindle and
speed and read the value from the display. [See Section II: Getting Started]
• The Brookfield PG Flash Software can be used to program the DV2T to control all
aspects of the test and data collection without the need for the operator to monitor
the instrument; just start the program and return to the printed test data (printer is
optional). [See Section V: PG Flash Software]
• The Brookfield RheocalcT Software will perform all control and data collection functions
of the DV2T from the PC while also providing a platform for advanced data collection
and analysis.
In any of these modes of control, the DV2T will provide the best in viscosity measurement
and control.
The principal of operation of the DV2T is to drive a spindle (which is immersed in the test fluid)
through a calibrated spring. The viscous drag of the fluid against the spindle is measured by the
Brookfield Dv2t Viscometer Manual Guide
spring deflection. Spring deflection is measured with a rotary transducer. The measurement
range of a DV2T (in centipoise or milliPascal•seconds) is determined by the rotational speed
of the spindle, the size and shape of the spindle, the container the spindle is rotating in, and
the full scale torque of the calibrated spring.
There are four basic spring torque series offered by Brookfield:
Spring Torque
Model
dyne
•cm
milliNewton
•m
DV2TLV
673.7 0.0673
DV2TRV
7,187.0
0.7187
DV2THA
14,374.0
1.4374
DV2THB
57,496.0
5.7496
The higher the torque calibration, the higher the measurement range. The measurement
range for each torque calibration may be found in Appendix B.